Abstract

AbstractIn recent years, the remote center of motion (RCM) constraint has moved from purely mechanical to software implementation, enabling the use of serial manipulators in robotic‐assisted minimal invasive surgery. However, ensuring safety with software‐based RCM presents challenges. This article addresses this issue by introducing a novel control algorithm for a 7‐DOF redundant robotic arm, taking into account a software RCM while considering system passivity and kinematic constraints. The algorithm is both a control optimizer and a robot kinematics inversion, enabling precise and dexterous control for various surgical tasks and other control applications. By formulating a quadratic optimization problem under linear constraints, the algorithm guarantees the robotic arm's performance, safety, and stability under communication delay. Experimental validation demonstrates the effectiveness and accuracy of the control algorithm in executing a surgical training task (peg‐and‐ring) during bilateral teleoperation. The results highlight the successful implementation of the RCM constraint, ensuring patient safety and optimizing the manipulation capabilities of the robotic arm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.