Abstract

A novel gaseous detector, based on parallel-plate gas amplification with a resistive anode and remote readout electrode, is described. The detector is significantly less prone to discharges than other gaseous wireless detectors and unique in its ability to achieve high gain at high counting rates. A local counting rate >5/spl times/10/sup 5/ counts/mm/sup 2/-s at a gas gain of up to 10/sup 5/ has been achieved in a 14/spl times/14 cm/sup 2/ area detector for diffraction applications. The detector radiation hardness is characterized by the anode accumulated charge >100 Coul/cm/sup 2/. The detector is sealed and its regassing period is estimated to be 5-10 yr. The operating characteristics of the new detector, its spatial resolution, and linearity are described and preliminary X-ray diffraction data are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.