Abstract
Polymethylpentene (PMP) oxygenators, utilised for ECMO, are commonly believed to be resistant to plasma leakage. Whilst uncommon, plasma leakage has been previously reported with PMP fibres, both in vivo and in vitro. We describe a paediatric ECMO case during which plasma leakage occurred and oxygenator function gradually deteriorated, ultimately necessitating device replacement. To our knowledge, this is the first case of plasma leakage described using a PMP device during paediatric ECMO. Subsequent investigation is described, demonstrating that a protein coating reduces the free passage of solution across the PMP membrane.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.