Abstract
We evaluated whether improved increase delivery efficiency of weight optimized dynamic conformal arc (WO-DCA) therapy in comparison to volumetric modulated arc therapy (VMAT) with single isocenter for SRS treatment of very small volume and multiple brain metastases (BMs). 20 patients having a less than 1 cc volume and 2≤, ≥4 of multiple BMs, redesigned for 20 Gy in 1 fraction using WO-DCA and VMAT techniques with double full coplanar and three partial noncoplanar arcs. Plan qualities were compared using tumor coverage, conformity index (CI), gradient index (GI), V4Gy, V10Gy, and V12Gy volumes of brain, monitor units (MUs), and percent of quality assurance pass rate (QA%). Both techniques satisfied clinical requirements in coverage and CI. VMAT had a significantly higher MU and mean GI than WO-DCA (for MUs; 2330 vs. 1991; P < 0.001, and for GI; 4.72 vs. 3.39; P < 0.001). WO-DCA was found significantly lower V4Gy (171.11 vs. 232.80 cm3, P < 0.001), V10Gy (25.82 vs. 29.71 cm3, P < 0.05), and V12Gy (14.35 vs. 17.28 cm3, P < 0.05) volumes than VMAT. WO-DCA was associated with markedly increase QA pass rates for all plans (97.65% vs. 92.64%, P < 0.001). WO-DCA may be the first choice compared to the VMAT in reducing the dose in the brain and minimizing small-field dosimetric errors for very small SRS treatment of brain metastases in the range of ≤ 1 cc and 2≤, ≥4.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.