Abstract

A novel vital-sign sensor with a self-injection-locked oscillator and a frequency demodulator to reduce system complexity and improve sensitivity is proposed. The theory provides a delta-sigma model to account for the excellent signal-to-noise spectral density ratio in a parametric study of the sensitivity performance. Then, the experiments verify the sensing principle and the predicted performance. Accordingly, a prototype sensor with high sensitivity is realized for noncontact cardiopulmonary monitoring, achieving a long sensing distance without the need for a low-noise amplifier. The sensing distance can grow four times longer by doubling the operating frequency. Furthermore, the sensor using a swept frequency can eliminate the null detection points and the external radio interference. As an experimental result, the sensor, which is placed 4 m away from the subject, can reliably detect the heartbeat signal an the operating frequency of 3.6 GHz and an output power level of 0 dBm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.