Abstract

A calibration procedure for a perturbation two-port vector network analyzer is presented. It consists of a variable perturbation two-port placed between a device-under-test and a scalar network analyzer. A measured vector reflection coefficient is determined on the basis of amplitude-only measurements. A full correction of systematic errors is possible. The new principle was experimentally confirmed in the frequency band up to 14 GHz. The results enable prediction of an add-on for scalar analyzers, enabling vector measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.