Abstract

A selective, specific, and sensitive ultra-performance LC (UPLC) method was developed for determination of eszopiclone and its degradation products. The chromatographic separation was performed with a Waters ACQUITY UPLC system and BEH C18 column using gradient elution with mobile phases A and B. Mobile phase A was 0.01 M phosphate buffer with 0.2% (w/v) 1-octane sulfonic acid sodium salt as an ion pair reagent, adjusted pH 2.2 with orthophosphoric acid-acetonitrile (85 + 15, v/v). Mobile phase B was pH 2.2 buffer-acetonitrile (20 + 80, v/v). UV detection was performed at 303 nm. Eszopiclone and its impurities were chromatographed with a total run time of 13 min. A calibration study showed that the response for each of the impurities A, B, C, and D was linear between concentrations of 0.02 and 7.2 microg/mL (r2 > or = 0.999). The method was validated over this range for precision, intermediate precision, accuracy, linearity, and specificity. For the precision study, RSD of each impurity was <5% (n = 6). The method was found to be precise, accurate, linear, and specific. The proposed method was successfully used for determination of eszopiclone impurities in pharmaceutical preparations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.