Abstract
A simple isocratic liquid chromatographic method was developed for determination of lopinavir from its related impurities and assay for the first time. This method involves the use of a C(8) (Symmetry Shield RP8, 150 x 4.6 mm, 5 microm) column. The method was validated over the range of limit of quantitation (LOQ) to 120% of impurity specification limit and LOQ to 150% of working concentration for assay. The mobile phase consisted of a mixture of 50 mM of potassium phosphate buffer, acetonitrile and methanol in the ratio of 40:50:10. The flow rate was set at 1.0 mL/min with UV detection monitored at 210 nm. The drug was subjected to stress conditions of hydrolysis, oxidation, photolysis and thermal degradation. The developed method was validated for linearity, range, precision, accuracy and specificity. This method was successfully applied for content determination of lopinavir in pharmaceutical formulations. The method can be conveniently used in a quality control laboratory for routine analysis for assay and related substances as well for the evaluation of stability samples of bulk drugs and pharmaceutical formulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.