Abstract

In this work, an ultrasensitive electrochemiluminescence (ECL) biosensor was constructed using poly-L-lysine (PLL) as a novel co-reactant of luminol and poly(luminol/aniline) nanorods loaded reduced graphene oxide (PLA@rGO) as nanoprobe, which enable highly sensitivity detection of glutathione (GSH). To the best of our knowledge, it is the first time that PLL was used for the co-reactant of luminol. Notably, about a 5-fold enhancement was obtained compared with the individual PLA@rGO using GCE. Due to the remarkable quenching effect between the excited state of PLL and the reduced form of GSH in the ECL system of luminol/PLL, the ECL sensing platform exhibited wide linear ranges of 1.0 × 10−9–1.0 × 10−4 M and 1.0 × 10−4–1.0 × 10−2 M and a low detection limit of 7.7 × 10−10 M. Simultaneously, the biosensor was also successfully applied to detect GSH in human serum sample with high recoveries. Hence, this work would open a new platform for the wide application of PLL in immunoassay and various sensors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.