Abstract

Biliary extraction baskets are a commonly used instrument for the removal of choledocholithiasis in endoscopic retrograde cholangiopancreatography (ERCP). Impaction of the extraction basket is a recognized complication of ERCP, and is usually the result of discrepancy between the size of bile duct stone and the diameter of the distal bile duct. Whilst mechanical lithotriptors can be used to crush the stone or break the wires of the basket to allow its release, failure of the lithotriptor device can occur. We describe the case of a 59-year-old gentleman who had an ERCP performed for choledocholithiasis. Basket impaction was encountered, and the mechanical lithotriptor failed to dislodge the stone/basket complex. A two-step technique involving balloon dilatation and forceps manipulation of the basket was applied to successfully dislodge the impacted basket. We believe this simple and safe technique should be adopted to rescue impacted biliary extraction baskets to avoid the need for potential surgical removal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.