Abstract
Conventional lasers ablate tissue through photothermal, photomechanical, and/or photoionizing effects, which may result in collateral tissue damage. The novel nonionizing picosecond infrared laser (PIRL) selectively energizes tissue water molecules using ultrafast pulses to drive ablation on timescales faster than energy transport to minimize collateral damage to adjacent cells. Animal cadaver study. Cuts in porcine laryngeal epithelium, lamina propria, and cartilage were made using PIRL and carbon dioxide (CO₂) laser. Lateral damage zones and cutting gaps were histologically compared. The mean widths of epithelial (8.5 μm), subepithelial (10.9 μm), and cartilage damage zones (8.1 μm) were significantly lower for cuts made by PIRL compared with CO₂ laser (p < 0.001). Mean cutting gaps in vocal fold (174.7 μm) and epiglottic cartilage (56.3 μm) were significantly narrower for cuts made by PIRL compared with CO₂ laser (P < 0.01, P < 0.05). PIRL ablation demonstrates superiority over CO₂ laser in cutting precision with less collateral tissue damage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.