Abstract

This paper presents a decoupling theory named as Vascular Loading Decoupling Technique (VLDT), extended from the beat-based Tissue Control Method (TCM), for noninvasive measurement of real-time-based intra-arterial blood pressure and dynamic compliance of blood vessel. Both VLDT and TCM are based upon the decoupling theory to cause the arterial pulsation without the influence from surrounding tissues and measure the variations of the vascular diameter at critical depth. Meanwhile, the AC part of blood pressure is lost, that is, the reference pressure for AC controller is absent as well. To remedy this problem, VLDT employs Step-Hold control rules and cubic spline curve fitting technique to estimate the reference pressure, identify the impedance of blood vessel, and calculate the AC control gain at Hold stage, then track the AC part of blood pressure and compute the real-time arterial blood pressure at Step stage by turns, which makes the real time measurement feasible; moreover, the dynamic compliance of blood vessel can be extracted from real-time impedances of blood vessel. Since the compliance of blood vessel is decoupled from surrounding tissues, it is valuable for diagnosing the severity of vascular sclerosis. The simulation results showed that the VLDT approach is superiority over TCM. This could provide new prospective to circulatory medicine research.

Highlights

  • This paper presents a decoupling theory named as Vascular Loading Decoupling Technique (VLDT), extended from the beat-based Tissue Control Method (TCM), for noninvasive measurement of real-time-based intra-arterial blood pressure and dynamic compliance of blood vessel

  • VLDT employs Step-Hold control rules and cubic spline curve fitting technique to estimate the reference pressure, identify the impedance of blood vessel, and calculate the AC control gain at Hold stage, track the AC part of blood pressure and compute the real-time arterial blood pressure at Step stage by turns, which makes the real time measurement feasible; the dynamic compliance of blood vessel can be extracted from real-time impedances of blood vessel

  • For obtaining the situation of (10), we present the real-time-based Vascular Loading Decoupling Technique (VLDT) instead of the beat-based Tissue Control Method (TCM) to detect the dynamic impedance of blood vessel and the instantaneous intra-arterial blood pressure

Read more

Summary

A Novel Theory

Measuring the Real-Time-Based Arterial Properties via Vascular Loading Decoupling Technique. How to cite this paper: Lin, H.C.H. and Lin, A.C.-Y. (2018) A Novel Theory: Measuring the Real-Time-Based Arterial Properties via Vascular Loading Decoupling Technique. Received: September 26, 2018 Accepted: November 27, 2018 Published: November 30, 2018

Introduction
Mechanical Circuit Model for Transmission of Arterial Blood Pressure
Analysis of Pulsation of Tissue and Artery
Vascular Loading Decoupling Technique
Creating the Decoupling Situation by Controlling the Arterial Blood Pressure
System Response Due to the Excitation of DC Part of Blood Pressure
System Response Due to the Excitation of AC Part of Blood Pressure
Simulation and Validation
Estimating the Reference Pressure by Featured Spline Curve Fitting Technique
Computing Impedance of Blood Vessel and Intra-Arterial Blood Pressure
Comparison the Simulation Results between VLDT and TCM
Findings
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.