Abstract
This study aimed to assess the feasibility, efficacy, and safety of a novel percutaneous postprocedure closure technique for large arterial sheath removal with the use of two Perclose ProGlide (Abbott Vascular Devices, Redwood City, CA) devices. Postprocedural closing of large-bore arteriotomies using the Perclose system can be difficult given the subsequent inability of the device to capture sufficient wall tissue. Our study was a single-center retrospective analysis of 22 consecutive patients who underwent large arteriotomy closure via the postclosure technique with a 12-16-Fr sheath. Efficacy endpoints included successful deployment of the system and hemostasis. Safety endpoints included the incidence of major or minor vascular complications as defined by the Vascular Academic Research Consortium-2 (VARC-2) definitions at 30-day follow-up. The postclosure technique resulted in 100% technical success rate and no postprocedural bleeding or vascular complications. Postclosure technique is a safe, highly effective, and feasible percutaneous method to achieve large-bore arteriotomy hemostasis with low rates of major bleeding or vascular complications and favorable early outcome.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.