Abstract

Abstract Introduction: In radiotherapy, the dose delivered outside the field is known as peripheral dose (PD). In this study, we have attempted to develop a dataset using the PD values measured with a two-dimensional array, IMatrixx. Methods: The IMatrixx was used to measure the PD up to a distance of 45 cm from the field edge, in a Varian Clinac 2100-C machine. Solid water slabs and water phantom were used to get the required geometry for the PD measurements. The measurements were done for different field sizes, collimator angles, source to surface distance (SSD) and depths. The influence of gantry angles and photon energies on the PD was studied. The surface dose measurements were carried out using thermoluminescent detectors (TLD). Results: The dataset shows that the PD increased significantly with field size and depth and its increase was insignificant for collimator rotation and SSD. The influence of gantry angle was less pronounced at dmax than at the surface. The TLD measurements at the surface of patients were in agreement with the IMatrixx measurements. Conclusions: The IMatrixx can be used for the generation of PD values and it is less time-consuming, accurate, and commonly available in all radiotherapy departments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call