Abstract

A new method for improving REM counter performance in Pulsed Neutron Fields (PNFs) has been developed. This method uses an analysis of the build-up of space charge in the counter to compensate for an underestimation of Ambient Dose Equivalent (H*(10)) in intense pulsed fields. It was applied to three sets of experimental data acquired using the LUPIN-II REM counter device, which is designed for use in PNFs. The data was acquired using the cyclotron at Helmholtz-Zentrum Berlin für Materialien und Energie GmbH (HZB), at the HiRadMat facility at CERN and at the ‘Elettra Sincrotrone Trieste’ (ELETTRA), Italy. A comparison of the data with and without this compensation method is used to highlight its effectiveness. The LUPIN-II performance, which has already been shown to be able to cope with fields of up to hundreds of nSv/burst, is improved by at least one order of magnitude, with further potential for improvement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.