Abstract
The main objective of this study was to develop a 3-D X-ray reconstruction system of the spine and rib cage for an accurate 3-D clinical assessment of spinal deformities. The system currently used at Sainte-Justine Hospital in Montreal is based on an implicit calibration technique based on a direct linear transform (DLT), using a sufficiently large rigid object incorporated in the positioning apparatus to locate any anatomical structure to be reconstructed within its bounds. During the time lapse between the two successive X-ray acquisitions required for the 3-D reconstruction, involuntary patient motion introduce errors due to the incorrect epipolar geometry inferred from the stationary object. An approach using a new calibration jacket and explicit calibration algorithm is proposed in this paper. This approach yields accurate results and compensates for involuntary motion occurring between X-ray exposures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.