Abstract

BackgroundPalpation of tumors during thoracoscopic surgery remains difficult, and identification of deep-seated tumors may be impossible. This preclinical study investigated the usefulness of a novel indocyanine green (ICG) fluorescence spectroscopy system for tumor localization.MethodsICG was diluted to 5.0×10−2 mg/mL in fetal bovine serum (FBS) and mixed with silicone resin to prepare pseudo-tumors. Sponges of different densities and a porcine lung were placed on top of the pseudo-tumors, which were examined using a novel fluorescence spectroscopy system and a near-infrared (NIR) camera. Spectra were measured for different sponge and lung thicknesses, and the lung spectra were measured during both inflation and deflation.ResultsThe fluorescence spectroscopy system was able to identify tumors at depths ≥15 mm, while the NIR system was not. The spectroscopy system also detected tumors at greater depths when the density of the intervening material was lower. Depending on the density and thickness of the intervening material, the system could detect spectra as deep as 40 mm for sponges and 30 mm for lungs.ConclusionsThis new fluorescence spectroscopy system can be used to identify lung tumors up to a depth of 30 mm in experiments using pseudo-tumors and a porcine lung, which may aid in tumor identification during thoracoscopic surgery.