Abstract

Due to recent technical advances in cataract surgeries, there has been a significant improvement in the safety and surgical outcomes of phacoemulsification. However, the corneal endothelium can be damaged during phacoemulsification by multiple factors. Therefore, we used a slit lamp to analyze the fluid dynamics of ophthalmic viscosurgical devices (OVDs) in the anterior chamber during phacoemulsification. In this experimental study, extracted porcine eyes were injected with OVDs stained with fluorescein through a side port of the eye and then fixed on a slit lamp microscope. After inserting a phaco tip, phacoemulsification simulation was then performed on the iris plane. Subsequent movements of OVDs in the anterior chamber were observed during the procedure by using a slit lamp microscope. Aspiration and removal of cohesive OVDs from the inside of the anterior chamber occurred within a few seconds after the ultrasonic vibration. Aspiration of dispersive OVDs occurred gradually, with some of the OVDs remaining on the side of the anterior chamber side in an irregular shape. This shape enabled the OVD to trap the air, thereby preventing the air from directly touching the corneal endothelium. Viscoadaptive OVDs remained inside the anterior chamber as a lump, with the infusion solution flowing between the corneal endothelium and the OVD, thus leading to the eventual aspiration of the OVD. Viscous dispersive OVDs remained as a lump between the corneal endothelium and the phaco tip. However, once the infusion solution flowed between the cornea and the OVD, the OVD detached from the corneal endothelium, indicating that this type would likely be aspirated and removed. This method, termed the “slit side view,” enables viewing of the movement of OVDs during surgery, as well as observation of the fluid dynamics in the anterior chamber.

Highlights

  • Today, most cataract surgeries are performed by using the phacoemulsification technique

  • The anterior chamber space can be observed in detail. erefore, we developed an observation method that uses a slit lamp microscope to obtain a detailed understanding of the movements that occur inside the anterior chamber during surgery. e purpose of this study was to evaluate the effectiveness of using a slit lamp to observe fluid dynamics in the anterior chamber during phacoemulsification

  • E ophthalmic viscosurgical devices (OVDs) studied was the viscoadaptive type. is type of OVD remained inside the anterior chamber as a lump, with the infusion solution flowing between the corneal endothelium and the OVD (Figure 6)

Read more

Summary

Research Article

A Novel “Slit Side View” Method to Evaluate Fluid Dynamics during Phacoemulsification. Erefore, we used a slit lamp to analyze the fluid dynamics of ophthalmic viscosurgical devices (OVDs) in the anterior chamber during phacoemulsification. In this experimental study, extracted porcine eyes were injected with OVDs stained with fluorescein through a side port of the eye and fixed on a slit lamp microscope. Viscoadaptive OVDs remained inside the anterior chamber as a lump, with the infusion solution flowing between the corneal endothelium and the OVD, leading to the eventual aspiration of the OVD. Viscous dispersive OVDs remained as a lump between the corneal endothelium and the phaco tip. Is method, termed the “slit side view,” enables viewing of the movement of OVDs during surgery, as well as observation of the fluid dynamics in the anterior chamber Once the infusion solution flowed between the cornea and the OVD, the OVD detached from the corneal endothelium, indicating that this type would likely be aspirated and removed. is method, termed the “slit side view,” enables viewing of the movement of OVDs during surgery, as well as observation of the fluid dynamics in the anterior chamber

Introduction
Journal of Ophthalmology
Phacoemulsification was performed in porcine eyes with
Discussion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.