Abstract

Measurement of the vascular access flow rate (Q(a)) is a widely accepted method for surveillance and predicting access failure. Among current practical methods, the ultrasound dilution technique is standard, but this requires a costly device available in few hemodialysis (HD) centers. Here, we devised a simple hemoglobin dilution technique to accurately measure Q(a) without the need for any special machines. Before HD, values of Q(a) were determined in each of 30 patients by hemoglobin dilution and then, in the same session, by ultrasound dilution. There was a significant correlation between the two techniques using automated hemoglobin and hematocrit or centrifuge-measured hematocrit levels to calculate HD fluid-derived Q(a) values. Our study shows that the HD dilution technique, using no special device, is economical, highly accurate, and easy to perform, and can be used as an alternative to standard ultrasound dilution for vascular access surveillance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.