Abstract

The authors have developed a non-invasive ultra- sound theragnostic system (NIUTS) with a focal lesion servo (FLS) function based on stereo ultrasound imaging in order to compensate for movement while irradiating high-intensity focused ultrasound (HIFU) onto a focal lesion. In the present paper, we propose a redundant motion control mechanism of the HIFU focus, for therapeutics, that is independent of the ultrasound probes for diagnostics. Using the proposed redundant motion control mechanism, an arbitrary designated position in the focal lesion can be treated without altering the viewpoint of the ultrasound probes. The proposed mechanism enables (i) noise factors, which deteriorate the image quality (IQ) for FLS, to be reduced, thereby enhancing the FLS performance, and (ii) enables the destruction of a focal lesion (kidney stone) with a preoperatively designated locus that is moving due to respiration/heartbeat.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.