Abstract
ReAces is a novel puncturable transcatheter atrial septal defect occluder. The device success rate was 100% (n=14 of 14) in the animal experiment. Four swine successfully received puncture of the device 60days after implantation of the device. The acute procedure success rate in the 10 patients was 100%. Transthoracic echocardiography examination showed that the devices were well positioned with no residual shunt, and the area of central portion of occluder was substantially thin. It is safe and effective to use ReAces for closure of secundum atrial septal defect, and puncture the atrial septum at the portion of the device is feasible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.