Abstract

We investigated the preclinical safety and efficacy of ventricular pulsed field ablation (PFA) using a family of novel, 6-/8-Fr, linear, and spiral PFA/mapping catheters (CRC EP, Inc). QRS-gated, bipolar PFA (>2.0 kV) was performed in 10 healthy swine. Altogether, 20 endocardial and epicardial right and left ventricular applications were delivered. The catheters were inserted through an 8.5-Fr steerable introducer. The intensity of skeletal muscle activation was quantified using an accelerometer. Lesions were assessed by pre- versus post-PFA electrogram analysis, pacing threshold, 3D voltage mapping, necropsy, and histology. The swine rete mirabile, liver and kidneys were examined for embolic events. All applications were single-shot (56 ± 18 s) without catheter repositioning. Minimal microbubbling was observed without significant skeletal muscle stimulation (mean acceleration 0.05 m/s2) or ventricular tachyarrhythmias. There was significant reduction in post- versus pre-PFA electrogram amplitude (0.5 ± 0.2 mV versus 3.2 ± 0.9 mV, P < 0.001) with a marked increase in pacing threshold (>20 mA versus 7.5 ± 2.9 mA, P < 0.001). All lesions were large and durable up to 28 days, measuring 32 ± 5 mm (length), 27 ± 8 mm (width), and 8 ± 3 mm (depth) using the spiral catheters and 43 ± 1 mm (length), 7 ± 1 mm (width), and 8 ± 1 mm (depth) using the linear catheters. Despite higher waveform voltages and prolonged applications, no thermal effects were detected at necropsy/histology. Moreover, gross and microscopic examinations revealed no evidence of thromboembolism, vascular or collateral injury. A novel, QRS-gated PFA system using linear and spiral PFA catheters is capable of creating large and durable ventricular lesions in vivo without significant microbubbling, ventricular arrhythmias or thromboembolism.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call