Abstract

Melamine selective acrylate citric acid (ACA) based polymeric membrane sensor was prepared by radical polymerization method and the sensor was characterized. The sensor showed a selective fluorescent response to melamine (λex/λem = 388/425 nm). The sensor response is linear in the concentration range of 3.96 × 10−9 to 7.93 × 10−8 mol L−1, the optimum pH value is 6.0 and response time is less than 1 min. Limit of detection (LOD) and limit of quantification (LOQ) were calculated as 2.32 × 10−10 mol L−1 and as 7.74 × 10−10 mol L−1, respectively. The sensor showed great selectivity towards melamine in the presence of a large number of molecules and ions. The performance of sensor was also confirmed by determining of melamine in milk powder sample and the results were compared with HPLC results and acceptable results were obtained. As a conclusion, the results revealed that the proposed sensor is an interesting alternative for melamine determination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.