Most ventricular assist devices (VADs) currently used in infants are extracorporeal. These VADs require long-term anticoagulation therapy and extensive surgery, and two devices are needed for biventricular support. We designed a biventricular assist device based on shape memory alloy that reproduces the hemodynamic effects of cardiomyoplasty, supporting the heart with a compressing movement, and evaluated its performance in a dedicated mockup system. Nitinol fibers are the device's key component. Ejection fraction (EF), cardiac output (CO), and generated systolic pressure were measured on a test bench. Our test bench settings were a preload range of 0-15 mm Hg, an afterload range of 0-160 mm Hg, and a heart rate (HR) of 20, 30, 40, and 60 beats/min. A power supply of 15 volts and 3.5 amperes was necessary. The EF range went from 34.4% to 1.2% as the afterload and HR increased, along with a CO from 180 to 6 ml/min. The device generated a maximal systolic pressure of 25 mm Hg. Cardiac compression for biventricular assistance in child-sized heart using shape memory alloy is technically feasible. Further testing remains necessary to assess this VAD's in vivo performance range and its reliability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.