Abstract

Magnetic localization provides a feasible way to estimate the position and orientation of the wireless capsule endoscopy embedded with a permanent magnet. However, the magnetic localization suffers from the drifting effect from magnetic noises and the uncertain initial guess during the optimization. In this paper, a localization system with noise cancellation is proposed to reduce the drifting effect of the geomagnetic noise and obtain the initial guess in a simple and effective way. Compared to the prior arts, the proposed system is able to achieve a lower positioning and orientation error under the same system configuration and makes it possible for the patient to move around while wearing the localization system. Besides, the proposed algorithm improves the consistency of positioning error within the localization region. Within a volume of 380 mm by 270 mm by 240 mm covered by 16 digital magnetic sensors, the average positioning error obtained from the proposed algorithm is around 10 mm and the average orientation error is around 12°.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.