Abstract

In this study, we assessed the diagnostic accuracy of the device VitaLog (SWG Sportwerk GmbH & Co. KG, Dortmund, Germany) for estimation of respiratory rate (RR) variability. VitaLog is a minimal-contact biomotion device that is placed under the mattress topper. It senses respiratory effort and body movement using a piezoelectric sensor. Diagnostic accuracy was determined in 103 patients referred to our sleep laboratory for suspected sleep-disordered breathing (SDB). SDB was defined by AHI ≥ 15/h. Results provided by VitaLog were compared with nasal flow measurement obtained by polysomnography (PSG). Diagnostic accuracy of VitaLog was excellent. We obtained a correlation of r = 0.99 and a bias of 0.2 cycles per minute (cpm) between VitaLog and PSG-provided nasal flow. Detection RR variability worked nearly identically in patients with and without SDB. VitaLog is an appropriate method for determination of RR variability based on a minimal-contact biomotion sensor. This device is easy to handle, available at low cost, and suitable for long-term monitoring in the hospital or at home.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.