Abstract

In this study, an optical system for simultaneous measurement of physical thickness, group refractive index, bow, and warp of a large silicon wafer is first proposed based on a reflection-type spectral-domain interferometer. Such key parameters are determined by combining four different optical path differences (OPDs) measured at each sampling point throughout two-axis sample scanning within area of 250 mm by 250 mm. To overcome the measurement limitations by the deflection of a free, unclamped large-sized wafer, two OPDs representing the surface profiles of both sides are utilized to facilitate the thickness and refractive index measurements insensitive to sample inclination. For verification of the proposed method, a 300 mm diameter silicon wafer with nominal thickness of 775 μm was used as a test sample. For measuring the bow and warp with gravity effect compensation, a silicon wafer was measured once again after turning over. Through theoretical analysis on the changes of OPDs with the wafer tilted based on the measured surface profiles, it was verified that the effect of wafer bending on thickness and refractive index measurements can be ignored. The measurement uncertainties (k = 1) of physical thickness, group refractive index, bow, and warp were evaluated to be approximately 0.692 μm, 0.003, 0.416 μm, and 0.589 μm, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.