Abstract
Wear remains an important cause of failure in knee replacement. Of the current methods of early performance assessment or prediction, simulators have been un-physiological, single X-ray film analyses remain limited by accuracy and retrieval and survival methods have a prohibitive time scale. An accurate method is needed to allow a timely assessment of polyethylene component wear in vivo, when a new design is introduced, in order to predict likely outcome. We present a new method for measuring wear in vivo that we believe will allow this prediction of long-term wear. X-ray film pairs were taken of implanted prosthetic metal components. When the X-ray system was calibrated, projections of the appropriate Computer Aided Design (CAD) model could be matched to the shapes on the scanned X-ray films to find component positions. Interpenetration of the metal femoral component into the polyethylene component could then be established and represents our estimate of “wear”. This method was used to measure in vivo prosthesis wear to an accuracy of 0.11 mm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.