Abstract
The primary and secondary prevention of sudden cardiac death resulting from malignant arrhythmia, channelopathy, and hypertrophic cardiomyopathy often requires the implantation of automatic internal cardiac defibrillators (AICDs) in the pediatric population. In young patients, the small size of the systemic veins, complex anatomy of congenital heart disease, and body habitus often preclude safe and durable transvenous placement of the AICD coil, requiring innovative methods to circumvent this problem. This report describes the technique used at Rady Children's Hospital San Diego/UCSD for the epicardial placement of an ICD system with a transvenous ICD coil placed between the aorta and pulmonary artery, thereby producing a stable location and excellent coil-to-can vector for successful defibrillation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.