Abstract
A novel electrolyte has been developed for the simultaneous separation of cations and anions in low explosive residue by capillary electrophoresis. This electrolyte contains 15 mM α-hydroxyisobutyric acid (HIBA) as the buffer, 6 mM imidazole as the cation chromophore, 3 mM 1,3,6-naphthalenetrisulfonic acid (NTS) as the anion chromophore, 4 mM 18-crown-6 ether as a cation selectivity modifier, and 5% (v/v) acetonitrile as an organic modifier. The pH was adjusted to 6.5 using tetramethylammonium hydroxide (TMAOH), an electroosmotic flow modifier. The method was optimized by varying the concentrations of α-HIBA, imidazole, and 1,3,6-NTS at three different pH values. The results provided a simultaneous indirect photometric analysis of both anions and cations with detection limits ranging from 0.5 to 5 ppm for anions and from 10 to 15 ppm for cations with a total run time of under 7 min. The method was then applied to the analysis of Pyrodex ® RS and black powder, as well as several smokeless powders. The results obtained were consistent with previously reported results for separate anion and cation analysis and provide a faster, more complete analysis of each sample in a single chromatographic run.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.