Abstract

For investigating the effects of mechanical ventilation on the respiratory system, experiments in small mammal models are used. However, conventional ventilators for small animals are usually limited to a specific ventilation mode, and in particular to passive expiration. Here, we present a computer-controlled research ventilator for small animals which provides conventional mechanical ventilation as well as new type ventilation profiles. Typical profiles of conventional mechanical ventilation, as well as flow-controlled expiration and sinusoidal ventilation profiles can be generated with our new ventilator. Flow control during expiration reduced the expiratory peak flow rate by 73% and increased the mean airway pressure by up to 1 mbar compared with conventional ventilation without increasing peak pressure and end-expiratory pressure. Our new ventilator for small animals allows for the application of various ventilation profiles. We could analyse the effects of applying conventional ventilation profiles, pressure-controlled ventilation and volume-controlled ventilation, as well as the novel flow-controlled ventilation profile. This new approach enables studying the mechanical properties of the respiratory system with an increased freedom for choosing independent ventilation parameters.