Abstract

In recent years, master-slave vascular robots have been developed to address the problem of radiation exposure during vascular interventions for surgeons. However, the single visual feedback reduces surgeon immersion and transparency of the system. In this work, we have developed a haptic interface based on the magnetorheological fluid (MRF) on the master side. The haptic interface can provide passive feedback force with high force fidelity and low inertia. Additionally, the manipulation of the master device does not change the operating posture of traditional surgery, which allows the surgeon to better adapt to the robotic system. For the slave robot, the catheter and guidewire can be navigated simultaneously which allows the two degrees of action on the catheter and axial action of a guidewire. The resistance force of the catheter navigation is measured and reflected to the user through the master haptic interface. To verify the proposed master-slave robotic system, the evaluation experiments are carried out in vitro, and the effectiveness of the system was demonstrated experimentally.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.