Abstract

ABSTRACTIn this study, a novel inflatable belt-type clamp is introduced and its performance is verified. Finite element simulations are performed to compare the performance of three different aorta clamping systems. In every case, the aorta is modeled as a simple hollow cylinder made of linearly elastic material. For a traditional surgical clamp in which the jaws remain inclined to one another as they close around the aorta, the maximum normal stress within the aorta wall is found to be 806kPa. It is shown that the numerical results are in good qualitative agreement with the experimental results obtained using a pressure sensitive film. The simulation results for a modified clamp in which the jaws remain parallel during the clamping operation show that the maximum normal stress is reduced to 222kPa. However, two regions of maximum stress are induced within the aorta wall. Finally, the numerical results for a novel inflatable belt-type clamp show that the maximum normal stress is equal to approximately 221kPa. In contrast to the modified clamp, the stress is uniformly distributed around the perimeter of the aorta, and thus the risk of aortic dissection is significantly reduced.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.