Abstract

Background EUS-guided FNA (EUS-FNA) is an established technique for the cytologic diagnosis of pancreatic disease. Attempts to obtain adequate histologic specimens have yielded variable and mostly insufficient results. Objective To evaluate the safety, feasibility, and quality of histologic biopsy specimens obtained by using a new cryobiopsy probe and to compare them with standard EUS-FNA and (laparoscopic) trucut biopsy specimens of pancreatic tissue. Design Animal non-survival study. Intervention Eighty-four pancreatic biopsy specimens (12 per group) were obtained in 4 anesthetized pigs by using one of the following the 18-gauge flexible cryoprobe; a conventional, 19-gauge, EUS-FNA needle; or a rigid, trucut biopsy device (18 gauge). The latter, used in laparoscopic surgery, was considered as the criterion standard for obtaining histology specimens. Main Outcome Measurements Specimens were evaluated for artifacts and specimen quality by a blinded pathologist who used a 7-point Likert scale to assess histologic adequacy. Biopsy size and bleeding time after biopsy also were recorded. Results The new cryoprobe was equivalent to the rigid, trucut needle and superior ( P Limitations Animal model. Conclusion EUS-guided cryobiopsy was associated with better specimen quality for histologic analysis and a shorter bleeding time compared with a conventional 19-gauge FNA needle in the animal model. It is a promising new technique for histologic examination of pancreatic tissue.