Abstract

To investigate the biological effects of a novel endovenous scaffold in a porcine model. Petalo is a compliant venous scaffold implanted into the internal jugular veins of 12 healthy pigs. The pigs were sacrificed at one, two, three, and six months, respectively. Microscopic investigations were performed at two blinded laboratories. Neo-intima formation progressively covering up the stent metallic bars was observed. The inflammatory response of the venous wall showed a peak after three months by the implant, followed by marked reduction after six months. The device induced a significant ( p < 0.01) increase of the thickness respect to the control regions, but was comparable in sections obtained after three and six months. The implant of Petalo compliant venous scaffold in the venous wall of this porcine model is characterized by neointima formation and by an inflammatory reaction which tends to decrease after six months. Our data point against the induction of smooth muscle cells proliferation and migration as confirmed by electronic transmission microscopy analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.