Abstract
Accurate measurement of six-degrees-of-freedom in-vivo kinematics of the total hip arthroplasty (THA) is essential in gaining insights into in-vivo THA performance. The objective of this study was to validate a novel dual fluoroscopy imaging system (DFIS) for determination of the THA kinematics using both in-vitro and in-vivo approaches. The in-vitro validation utilized cadaveric hip specimens to compare the THA motion using the DFIS technique with those measured by a radiostereometric analysis (RSA). The differences between the DFIS technique and the RSA were within 0.33±0.81mm (mean±SD) in translation and 0.45±0.65° in rotation during dynamic motion of the hips. In the in-vivo validation, the THA kinematics of two patients during a treadmill gait was assessed for the feasibility/repeatability of the DFIS technique in measurement of THA kinematics. The poses of the THAs during the treadmill gait was measured using the DFIS technique with the maximum standard deviation of 0.35mm in translation and of 0.55° in rotation. This study demonstrated that the DFIS technique has comparable accuracy of the RSA and is highly repeatable for measurement of dynamic THA motion, suggesting that the DFIS is a promising tool in evaluating the in-vivo THA biomechanics during functional activities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.