Abstract

A simulated moving bed (SMB) chromatography system is a powerful tool for preparative scale separation, which can be applied to the separation of chiral compound. We have designed our own lab-scale SMB chromatography using 5 HPLC pumps, 6 stainless steel columns and 4 multi-position valves, to separate a racemic mixture of ketoprofen in to its enantiomers. Our design has the characteristics of the low cost for assembly for the SMB chromatography and easy repair of the unit, which differs from the designs suggested by other investigators. It is possible for the flow path through each column to be independently changed by computer control, using 4 multi-position rotary valves and 5 HPLC solvent delivery pumps. In order to prove the operability of our SMB system, attempts were made to separate the (S)-ketoprofen enantiomer from a ketoprofen racemic mixture. The operating parameters of the SMB chromatography were calculated for ketoprofen separation from a batch chromatography experiment as well as by the triangle theory. With a feed concentration of 1 mg/mL, (S)-ketoprofen was obtained with a purity of 96% under the calculated operating conditions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call