Abstract

The non-invasive fetal electrocardiogram (fECG) enables easy detection of developing heart abnormalities, leading to a significant reduction in infant mortality rate and post-natal complications. Due to the overlapping of maternal and fetal R-peaks, the low amplitude of the fECG, systematic and ambient noises, typical signal extraction methods, such as adaptive filters, independent component analysis, empirical mode decomposition, etc., are unable to produce satisfactory fECG. While some techniques can produce accurate QRS waves, they often ignore other important aspects of the ECG. Utilizing extensive preprocessing and an appropriate framework, our approach, built upon 1D CycleGAN, achieves fECG signal reconstruction from the mECG signal while preserving its morphology. The performance of our solution was evaluated by combining two available datasets from Physionet, “Abdominal and Direct Fetal ECG Database” and “Fetal electrocardiograms, direct and abdominal with reference heartbeat annotations”, where it achieved an average PCC and Spectral-Correlation score of 88.4% and 89.4%, respectively. It detects the fQRS of the signal with accuracy, precision, recall and F1 score of 92.6%, 97.6%, 94.8% and 96.4%, respectively. It can also accurately produce the estimation of fetal heart rate and R-R interval with an error of 0.25% and 0.27%, respectively. The main contribution of our work is that, unlike similar studies, it can retain the morphology of the ECG signal with high fidelity. The accuracy of our solution for fetal heart rate and R-R interval length is comparable to existing state-of-the-art techniques. This makes it a highly effective tool for early diagnosis of fetal heart diseases and regular health checkups of the fetus.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call