Abstract
In ultrasound color flow imaging systems, it is important to suppress the clutter signals originated from stationary and slowly moving tissue sufficiently. Without sufficient clutter rejection, the estimation of the blood flow velocity will be inaccurate, and the imaging quality will be poor. In this paper we present a new clutter rejection scheme involving first down-mixing the clutter signals to zero frequency and then canceling the stationary echo by subtracting the signals from two consecutive pulses. This stationary echo canceling scheme is easy to be realized with small computational power, and is adaptive to the clutter signals. Theoretical analysis and simulation using RF data prove that it’s an efficient and practical clutter rejection scheme for color flow imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.