Abstract
Peripheral neuropathy with antibodies to myelin-associated glycoprotein (MAG) is an autoimmune demyelinating disorder of the peripheral nervous system caused by pathogenic IgM recognizing the human natural killer-1 glycoepitope expressed on MAG. This study aimed to analyze the performance of a new indirect immunofluorescence cell-based assay (CBA, EUROIMMUN) for the detection of anti-MAG IgM. Antibody reactivity was determined in sera from 95 patients with clinical and neurophysiological evidence of anti-MAG-associated neuropathy and in control samples from 55 patients with other forms of peripheral neuropathy. Compared to the results of the gold standard method (ELISA, Bühlmann) and using samples at a dilution of 1:100, the CBA had a sensitivity of 98.9% and a specificity of 100% (PPV 100%, NPV 98.2%). In conclusion, the CBA allows the detection of antibodies to MAG using an easy and standardized technique, and it presents a sensitive and specific alternative to the more time-consuming ELISA. Larger studies are needed to address anti-MAG titer monitoring in parallel with clinical activity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.