Abstract

ABSTRACT A novel SPECT model with an elliptical orbit and multiple detectors is proposed for detecting the metabolism of the human heart. Under the structure of the proposed model, the distance between the detector and the patient becomes smaller, which can effectively improve the SNR. The slit-slat collimator is used to collimate the emitted photons. Modified iterative reconstruction algorithm is proposed to reconstruct a small centre field with severe truncation data under the constraints of body contour and total variation (TV). Simulations for the proposed SPECT are implemented to access the image reconstruction quality. The simulation results show that proposed imaging system with appropriate reconstruction methods can provide significant improvements in the spatial resolution and the image quality. The geometric efficiency of the proposed SPECT is about twice that of the conventional SPECT system, but the scanning time is only about one tenth.