Abstract

Two non-electroactive biomarkers methylglyoxal (MGo) and glyoxal (Go) in urine and environmental water samples were determined for the first time by capillary electrophoresis with amperometric detection (CE-AD) after derivatizing with an electroactive compound 2-thiobarbituric acid. Experimental conditions of derivatization and CE-AD detection were optimized. Highly linear response was obtained for these two biomarkers over three orders of magnitude with good correlation ( r 2 > 0.999). The limits of detection (LODs) and limits of quantitation (LOQs) of MGo and Go were 0.2 μg L −1 and 1.0 μg L −1, 0.5 μg L −1 and 2.0 μg L −1, respectively. The average recovery and relative standard deviation (RSD) were within the range of 90.9–101.3% and 0.7–2.2%, respectively. The proposed CE-AD method provides a reliable and sensitive quantitative evaluation of MGo and Go in real sample matrices by employing relatively simple and inexpensive instrument.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.