Abstract
Placement of intraventricular catheters in oncology patients can be associated with morbidity given their small to slit-like ventricles and underlying hematologic disorders. We studied the accuracy of placing Ommaya reservoirs using neuronavigation and a flexible neuroendoscope to verify catheter positioning. Ommaya reservoirs placed in 25 oncology patients between 2013 and 2015 were retrospectively reviewed. Twenty-five ventricular catheters were placed using the AxiEM stealth frameless neuronavigation system and a flexible neuroendoscope. Postoperative catheter accuracy, operative complications, and postoperative complications were assessed. We discuss surgical protocol and technical nuances. All ventricular catheters were successfully placed into the ipsilateral (84%) or contralateral (16%) foramen of Monro. A single ventricular catheter pass was needed to cannulate the ventricle in 96% of patients. The mean accuracy was 4.09 ± 3.47 mm from the target, the ipsilateral foramen of Monro. One patient had a catheter tract hemorrhage seen on postoperative imaging related to thrombocytopenia. No postoperative neurologic deficits were seen. A combined neuronavigation and neuroendoscopic approach improved catheter tip accuracy compared with accuracy rates described in the literature using other techniques. This approach can be adapted toward routine clinical practice of placing ventricular shunt catheters and Ommaya reservoirs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.