Abstract

Four-channel flow-through electrochemical cell working in thin-layer regime was designed, fabricated and characterized experimentally and in computational fluid dynamics (CFD) simulations. The new principle of operation allows reproducible splitting of a stream of liquid into multiple flow channels. Systems comprising of 2-, 3-, 4- and 8-channels were tested. The proper function of the cell is given by the ratio of the cross-sections of the fluidic element collecting chamber and the particular flow paths among which the liquid is distributed. Suitable flow rates providing uniform liquid distribution were evaluated and the results were compared to CFD modeling. The flow-through cells designed according to the proposed principle can be simply incorporated in automated routine analysis as only one inlet and one common outlet are required.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.