Abstract

To demonstrate a novel approach to scleral fixation of posterior chamber intraocular lenses and capsular tension rings and segments in deep-set eyes using the Finesse FlexLoop (Alcon Laboratories). The technique described herein, based on previous approaches to scleral fixation of posterior chamber intraocular lenses, uniquely employs the FlexLoop to "lasso" Gore-Tex sutures that have already been threaded through the eyelets of a CZ70BD (Alcon Laboratories) IOL and externalize them. All patients who underwent surgery with this technique experienced visual improvement. The only complication was of mild hyphema in the patient who had a capsular tension segment placed, which resolved with medical therapy. The advantages of this procedure include a smaller diameter instrument (FlexLoop) as compared to the 25-gauge forceps typically employed, an easier to perform surgical maneuver that alleviates the need for both precise placement and constant tension to be exerted by the surgeon to grasp the sutures, as well as an instrument that can function when bent up to 45° to help accommodate deep-set eyes requiring this procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.