Abstract

Bacterial leakage from the implant-abutment-interface (IAI) is suspected of contributing to the development of peri-implantitis. The aim of the study was to develop a straightforward test setup to evaluate the bacterial leakage of the IAI of 2-piece implant systems under laboratory conditions. A test suspension of Enterococcus faecium was injected into 7 implants (PerioType Rapid Implants) prior to abutment fixation. The IAI was covered by kanamycin aesculn azide agar (KAAA), which serves as an optical indicator for E. faecium . The specimens were cyclically loaded with a force of 120 N for up to 1 000 000 cycles in a universal testing machine in accordance with the ISO 14801:2007 standard. The color change of the KAAA was recorded. Three of the 7 implants showed bacterial leakage before the cyclic loading test started. The bacterial tightness of the IAIs of the 4 remaining implants lasted for 35 680 ± 22 467 cycles on average. The experimental setup at hand provides the means for a straightforward evaluation of the bacterial tightness of the IAI of 2-piece dental implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.