Abstract
Although ex vivo simulation is a valuable tool for surgical optimization, a disease model that mimics human aortic regurgitation (AR) from cusp prolapse is needed to accurately examine valve biomechanics. To simulate AR, four porcine aortic valves were explanted, and the commissure between the two largest leaflets was detached and re-implanted 5mm lower to induce cusp prolapse. Four additional valves were tested in their native state as controls. All valves were tested in a heart simulator while hemodynamics, high-speed videography, and echocardiography data were collected. Our AR model successfully reproduced cusp prolapse with significant increase in regurgitant volume compared with that of the controls (23.2 ± 8.9 versus 2.8 ± 1.6ml, p = 0.017). Hemodynamics data confirmed the simulation of physiologic disease conditions. Echocardiography and color flow mapping demonstrated the presence of mild to moderate eccentric regurgitation in our AR model. This novel AR model has enormous potential in the evaluation of valve biomechanics and surgical repair techniques. Graphical Abstract.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.