Abstract
A new antibody overlay technique for 2-dimensional immunoelectrophoresis (2-D-IEP) is described. After the first dimension electrophoresis of the antigen, the desired amount of antiserum is applied to the initial agarose layer and then evenly distributed over a defined surface of the gel with a PVC film. This modification of the conventional 2-D-IEP procedure makes it possible to perform tandem 2-D-IEPs comparing 2 antisera on the same gel plate, rocket IEPs where several antisera are compared by electrophoresis on the same gel plate, and 2-D-IEPs with an intermediate antiserum, avoiding the need to pour an intermediate gel. With this technique, 77 antigens have been demonstrated in a Candida albicans serotype A somatic antigen preparation. This is the first description of such a large number of immunoprecipitates on the same immunoplate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.