An alternative liquid chromatography-mass spectrometry (LC-MS) method based on no discharge (ND) atmospheric pressure chemical ionization (APCI) was developed for the simultaneous determination of ethyl glucuronide (EtG) and ethyl sulfate (EtS) in urine in negative ion conditions. Abundant [M-H](-) species of EtG and EtS were obtained, allowing to reach limits of quantification (0.1 microg/ml for both analytes), accuracy, and precision comparable to those proposed in the literature. Additionally, the LC-ND-APCI-MS method proved to be reliable, requiring little maintenance even when high throughput analyses (i.e., 6,000 samples per year) were required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.