Abstract
This paper presents a novel 10GHz bio-radar system based on a frequency multiplier and phase-locked loop (PLL) for non-contact measurement of heartbeat and respiration rates. In this paper, a 2.5GHz voltage controlled oscillator (VCO) with PLL is employed as a frequency synthesizer, and 10GHz continuous wave (CW) signal is generated by using frequency multiplier from 2.5GHz signal. This paper also presents the noise characteristics of the proposed system, and the analysis result shows that the same signal-to-noise-ratio (SNR) performance can be achieved with the proposed system based on the frequency multiplier compared with the conventional system with identical carrier frequency. The experimental results shows excellent vital-signal measurement up to 100cm without any additional digital signal processing (DSP), thus proving the validity of the proposed system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.